4. Test Strip has not
been fully inserted
into the Meter.
4. Reinsert the Test
Strip into the Meter.
Test results
are incon
sistent or
Control
Solution test
results are
not within
the specified
range.
1. Not enough sample
in the Test Strip.
2. Test Strip or Control
Solution has expired.
3. Test Strip has been
damaged due to
heat or humidity so
that sample cannot
be applied or the
speed of applying
sample is too slow.
4. System is not per-
forming due to the
environment being
above or below
room temperature.
1. Redo test with new
test strip and make
sure that enough
sample has been
added.
2. Redo test with new
test strip.
3. Perform a Control
Solution test using a
new Test Strip. If
results are still out of
range, replace with
new vial of Test Strips.
4. Bring System to a
room temperature
environment and
wait approximately 30
minutes before per-
forming a new test.
The Meter co-
untdown did
not start.
Test Strip has not
been inserted
correctly.
Use a new Test Strip
and redo the test.