Important Control Solution Information
Use only CLEVER CHOICE control solutions with your meter.
Do not use the control solution beyond the expiration date or 3
months after rst opening. Write the opening date on the control
solution vial and discard the remaining solution after 3 months.
It is recommended that the control solution test should be done at
room temperature (68°F-77°F/20°C-25°C). Make sure your control
solution, meter, and test strips are at this specied temperature
range before testing.
Shake the vial before use, discard the rst drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an ac-
curate result.
Store the control solution tightly closed at temperatures between
36°F and 86°F (2°C to 30°C). Do NOT freeze.
Do not use test strips beyond the expiration date. This may cause
inaccurate results.
Do not bend, cut, or alter a test strip in any way.
Keep the strip vial away from children since the cap and the test
strip may be a choking hazard. If swallowed, promptly see a doctor
for help.
For further information, please refer to the test strip package insert.